OBJECTIVES: This study reports on the performance of a leadless ventricular pacemaker with automated, enhanced accelerometer-based algorithms that provide atrioventricular (AV) synchronous pacing.
BACKGROUND: Despite many advantages, leadless pacemakers are currently only capable of single-chamber ventricular pacing.
METHODS: The prospective MARVEL 2 (Micra Atrial tRacking using a Ventricular accELerometer 2) study assessed the performance of an automated, enhanced accelerometer-based algorithm downloaded to the Micra leadless pacemaker for up to 5 h in patients with AV block. The primary efficacy objective was to demonstrate the superiority of the algorithm to provide AV synchronous (VDD) pacing versus VVI-50 pacing in patients with sinus rhythm and complete AV block. The primary safety objective was to demonstrate that the algorithm did not result in pauses or heart rates of >100 beats/min.
RESULTS: Overall, 75 patients from 12 centers were enrolled; an accelerometer-based algorithm was downloaded to their leadless pacemakers. Among the 40 patients with sinus rhythm and complete AV block included in the primary efficacy objective analysis, the proportion of patients with ≥70% AV synchrony at rest was significantly greater with VDD pacing than with VVI pacing (95% vs. 0%; p
* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.