Affiliations 

  • 1 A Mohd Azlan, MS (Orth). Orthopaedic Unit, Department of Surgery, Universiti Putra Malaysia, Kuala Lumpur, Malaysia
  • 2 A R Mohammad, MS (Orth). Department of Othopaedics and Traumatology, Hospital UKM, Cheras, Kuala Lumpur
  • 3 A K Ariffin, PhD. Department of Mechanical and Materials Engineering, Universiti Kebangsaan Malaysia, Bangi, Selangor, Malaysia
Med J Malaysia, 2005 Jul;60 Suppl C:30-4.
PMID: 16381280 MyJurnal

Abstract

This finite element analysis is aimed at comparing relative stiffness of three different posterior instrumentation constructs: the Hospital Universiti Kebangsaan Malaysia Spinal Instrumentation System (HUKM-SIS), the Cotrell-Dubousset Instrumentation (CDI) and Harrington Instrumentation System (HIS), used in the treatment of adolescent idiopathic scoliosis (AIS). The constructs were tested under various loads using MSC Patran 2001 r2a. Under increasing flexion loads, there was a linearly corresponding increase in deflection magnitudes for all constructs on the load-deflection curve. The CDI was the stiffest construct under axial, forward flexion and extension loads, followed by the HUKM-SIS and HIS. Under lateral bending loads, the HUKM-SIS construct was the stiffest followed by CDI and HIS. The HUKM-SIS construct was stiffer than HIS under torsional loads. We conclude that multiple pedicle screws increase the stiffness of posterior instrumentation constructs under all loads and inter-segmental spinous processes wiring increase the stiffness against lateral bending.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.