Affiliations 

  • 1 Audiology Programme, Centre for Rehabilitation and Special Needs Studies, Universiti Kebangsaan Malaysia, Kuala Lumpur, Malaysia
  • 2 Faculty of Health Sciences, National Centre for Audiology, Western University, London, Canada
Int J Audiol, 2022 11;61(11):924-931.
PMID: 34859745 DOI: 10.1080/14992027.2021.2005832

Abstract

OBJECTIVE: This study examined the accuracy of hearing aid output to DSL v5.0 Child targets in a group of children who wear hearing aids, and the impact on their auditory outcomes.

DESIGN: For each participant, the output of the initial hearing aid fitting was compared to DSL v5.0 Child prescriptive targets and again after the fitting was adjusted using coupler-based verification and RECD measures. Outcomes for initial and adjusted fittings were examined using the Speech Intelligibility Index (SII), Parent's Evaluation of Aural/Oral Performance of Children (PEACH) rating scale, and speech perception tests in quiet and noise.

STUDY SAMPLE: Sixty-eight children aged 3 months to 17 years with moderate to profound hearing loss participated in the study.

RESULTS: Fit-to-targets improved significantly after hearing aids were adjusted to match targets to within 5 dB RMSE. Adjusted hearing aids provided increased aided audibility compared to initial fittings and resulted in improved speech perception scores and parent-reported hearing performance. Fifty percent of the children aged 6 to 17 years preferred their adjusted fitting compared to 10% who preferred their initial fitting.

CONCLUSIONS: Improvement in fit-to-target to a validated paediatric prescriptive formula using best practice procedures can result in improved auditory outcomes and possible self-reported satisfaction.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.