  1. Nazaimoon W, Ng ML, Bak K
    Malays J Pathol, 1993 Jun;15(1):75-83.
    PMID: 8277795
    A simple, non-isotopic in-house enzyme-linked immunoabsorbant assay (ELISA) for human growth hormone (GH) was developed. The assay involved using in-house polyclonal anti-GH adsorbed onto 96-well microtitre plates, commercially prepared mouse monoclonal anti-GH, and goat anti-mouse IgG horseradish peroxidase detection system. Results of recovery and parallelism studies ranged from 95%-106% and 98%-101% respectively, of the expected values. The detection limit of the assay was 0.008 mIU/well or the equivalent to 0.4 mIU/L of undiluted serum. Intra- and interassay coefficients of variations were 4.8%-7.9% and 6.5%-8.7% respectively. Serum GH levels measured in this assay correlated well with those measured in established in-house radioimmunoassays (r = 0.985, p < 0.001) and immunoradiometric assay from NETRIA (r = 0.984, p < 0.001).