Displaying all 3 publications

Abstract:
Sort:
  1. Rabba JA, Jaafar HA, Suhaimi FM, Jafri MZM, Osman ND
    J Med Radiat Sci, 2023 Nov 15.
    PMID: 37965811 DOI: 10.1002/jmrs.738
    INTRODUCTION: A standardised testing protocol for evaluation of a wide range of dental cone beam computed tomography (CBCT) performance and image quality (IQ) parameters is still limited and commercially available testing tool is unaffordable by some centres. This study aims to assess the performance of a low-cost fabricated phantom for image quality assessment (IQA) of digital CBCT unit.

    METHODS: A customised polymethyl methacrylate (PMMA) cylindrical phantom was developed for performance evaluation of Planmeca ProMax 3D Mid digital dental CBCT unit. The fabricated phantom consists of four different layers for testing specific IQ parameters such as CT number accuracy and uniformity, noise and CT number linearity. The phantom was scanned using common scanning protocols in clinical routine (90.0 kV, 8.0 mA and 13.6 s). In region-of-interest (ROI) analysis, the mean CT numbers (in Hounsfield unit, HU) and noise for water and air were determined and compared with the reference values (0 HU for water and -1000 HU for air). For linearity test, the correlation between the measured HU of different inserts with their density was studied.

    RESULTS: The average CT number were -994.1 HU and -2.4 HU, for air and water, respectively and the differences were within the recommended acceptable limit. The linearity test showed a strong positive correlation (R2  = 0.9693) between the measured HU and their densities.

    CONCLUSION: The fabricated IQ phantom serves as a simple and affordable testing tool for digital dental CBCT imaging.

  2. Rabba JA, Suhaimi FM, Mat Jafri MZ, Jaafar HA, Osman ND
    Radiography (Lond), 2023 May;29(3):533-538.
    PMID: 36913788 DOI: 10.1016/j.radi.2023.02.028
    INTRODUCTION: The daily image quality assessment involves large datasets that consume a lot of time and effort. This study aims to evaluate a proposed automated calculator for image distortion analysis in 2-dimensional (2D) panoramic imaging mode for a dental cone beam computed tomography (CBCT) system in comparison with present manual calculations.

    METHODS: A ball phantom was scanned using panoramic mode of the Planmeca ProMax 3D Mid CBCT unit (Planmeca, Helsinki, Finland) with standard exposure settings used in clinical practice (60 kV, 2 mA, and maximum FOV). An automated calculator algorithm was developed in MATLAB platform. Two parameters associated with panoramic image distortion such as balls diameter and distance between middle and tenth balls were measured. These automated measurements were compared with manual measurement using the Planmeca Romexis and ImageJ software.

    RESULTS: The findings showed smaller deviation in distance difference measurements by proposed automated calculator (ranged 3.83 mm) as compared to manual measurements (ranged 5.00 for Romexis and 5.12 mm for ImageJ software). There was a significant difference (p 

  3. Samson DO, Aziz MZA, Shukri A, Mat Jafri MZ, Hashim R, Zuber SH, et al.
    Health Phys, 2023 Aug 01;125(2):77-91.
    PMID: 36826380 DOI: 10.1097/HP.0000000000001688
    The current study was undertaken to investigate the radiological and dosimetric parameters of natural product-based composite (SPI/NaOH/IA-PAE/ Rhizophora spp .) phantoms. The radiological properties of the phantoms were measured at different gamma energies from Compton scatter of photons through angles of 0, 30, 45, 60, 75, and 90 degrees. Ionization chamber (IC) and Gafchromic EBT3 film dosimeters were employed to evaluate the dosimetric characteristics for photons (6-10 MV) and electrons (6-15 MeV). Radiological property results of the composite phantoms were consistent with good quality compared to those of solid water phantoms and theoretical values of water. Photon beam quality index of the SPI15 phantom with p-values of 0.071 and 0.073 exhibited insignificant changes. In addition, good agreement was found between PDD curves measured with IC and Gafchromic EBT3 film for both photons and electrons. The computed therapeutic and half-value depth ranges matched within the limits and are similar to those of water and solid water phantoms. Therefore, the radiological and dosimetric parameters of the studied composite phantom permit its use in the selection of convenient tissue- and water-equivalent phantom material for medical applications.