BACKGROUND: Leadless pacemakers may be preferred in patients with limited vascular access and high-infection risk, such as patients on hemodialysis.
METHODS: Patients on hemodialysis at the time of Micra implantation attempt (n = 201 of 2,819; 7%) from the Micra Transcatheter Pacing Study investigational device exemption trial, Micra Transcatheter Pacing System Continued Access Study Protocol, and Micra Transcatheter Pacing System Post-Approval Registry were included in the analysis. Baseline characteristics, periprocedural outcomes, and intermediate-term follow-up were summarized.
RESULTS: Patients on hemodialysis at the time of Micra implantation attempt were on average 70.5 ± 13.5 years of age and 59.2% were male. The dialysis patients commonly had hypertension (80%), diabetes (61%), coronary artery disease (39%), and congestive heart failure (27%), and 72% had a condition that the implanting physician felt precluded the use of a transvenous pacemaker. Micra was successfully implanted in 197 patients (98.0%). Reasons for unsuccessful implantation included inadequate thresholds (n = 2) and pericardial effusion (n = 2). The median implantation time was 27 min (interquartile range: 20 to 39 min). There were 3 procedure-related deaths: 1 due to metabolic acidosis following a prolonged procedure duration in a patient undergoing concomitant atrioventricular nodal ablation and 2 deaths occurred in patients who needed surgical repair after perforation. Average follow-up was 6.2 months (range 0 to 26.7 months). No patients had a device-related infection or required device removal because of bacteremia.
CONCLUSIONS: Leadless pacemakers represent an effective pacing option in this challenging patient population on chronic hemodialysis. The risk of infection appears low with an acceptable safety profile. (Micra Transcatheter Pacing Study; NCT02004873; Micra Transcatheter Pacing System Continued Access Study Protocol; NCT02488681; Micra Transcatheter Pacing System Post-Approval Registry; NCT02536118).
METHODS AND RESULTS: The primary outcome (risk of cardiac failure, pacemaker syndrome, or syncope related to the Micra system or procedure) was compared between successfully implanted patients from the Micra IDE trial with a primary pacing indication associated with AF or history of AF (AF group) and those without (non-AF group). Among 720 patients successfully implanted with Micra, 228 (31.7%) were in the non-AF group. Reasons for selecting VVI pacing in non-AF patients included an expectation for infrequent pacing (66.2%) and advanced age (27.2%). More patients in the non-AF group had a condition that precluded the use of a transvenous pacemaker (9.6% vs. 4.7%, P = 0.013). Atrial fibrillation patients programmed to VVI received significantly more ventricular pacing compared to non-AF patients (median 67.8% vs. 12.6%; P
OBJECTIVE: The purpose of this study was to report the performance of the Micra AV leadless pacemaker from the worldwide Micra AV post-approval registry (PAR) through 12 months.
METHODS: The Micra AV PAR is a prospective, single-arm, observational registry designed to assess the safety and effectiveness of Micra AV in a real-world setting. For the present interim analysis, major complications and system revisions through 12 months were summarized and compared to a historical cohort of 2667 patients implanted with a transvenous dual-chamber pacing system.
RESULTS: The device was successfully implanted in 796 of 801 patients (99.4%) at 97 centers in 19 countries. Micra AV patients were older (74.1 years vs 71.1 years; P < .0001) with a higher incidence of renal disease (22.3% vs 9.8%; P < .0001) compared with transvenous dual-chamber patients. Through 12 months, the major complication rate was 3.7% in Micra AV patients compared with 8.8% in transvenous dual-chamber patients (hazard ratio 0.42; 95% confidence interval 0.28-0.61; P < .001). The system revision rate was 1.5% in Micra AV patients compared with 5.5% for transvenous dual-chamber patients (hazard ratio 0.25; 95% confidence interval 0.13-0.47; P < .001); this reduction was largely driven by the absence of lead dislodgments requiring revision. The median AV synchrony index was 79.4% (interquartile range 65.2%-86.4%) in patients paced >90%.
CONCLUSION: The Micra AV leadless pacemaker was implanted with a high rate of success in patients with multiple comorbidities, with a significantly lower rate of complications and system revisions through 12 months compared to a historical cohort of patients with transvenous dual-chamber pacemakers.