Affiliations 

  • 1 Jabatan Sains Bioperubatan, Fakulti Sains Kesihatan Bersekutu, Universiti Kebangsaan Malaysia, Jalan Raja Muda Abdul Aziz, 50300 Kuala Lumpur
Malays J Med Sci, 1999 Jul;6(2):12-6.
PMID: 22589683 MyJurnal

Abstract

The development of antibody-based biosensor has grown steadily during recent years, and their use as a routine instrument in clinical application is not far from reality. This study has demonstrated the capability of conductometric sensor to quantitate human Follicle Stimulating Hormone (hFSH) from urine samples. The principles are adopted from Enzyme Linked Immunosorbent Assay (ELISA) technique. Self fabricated gold coated electrode was dipped in the microtiter well containing antibody-antigen complex. Substrate was added to the system to initiate a secondary reaction, which produced electroactive species and change the conductivity of the solution. The changes were proportional with the concentration of the hormone present. The results obtained correlate well with the conventional ELISA technique. Inter and intra assay variation (%CV) were under 6% and the lowest detection limit is 0.75 mIU/ml which was well under the physiological range of the hormone. This system offered advantages such as simplicity, reliability, minimal addition of reagents, freedom from turbidity and color problem, probability of miniaturizing the electrode thus minimizing the sample volume and the ability of on line data analysis. This study proved that Antigen-Antibody reaction via EIA could be detected electronically and it has a potential to be used as one of the measuring mode in clinical analysis.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.

Similar publications